Cl. Yang et al., THERMOSPRAY NEBULIZER AS SAMPLE INTRODUCTION TECHNIQUE FOR MICROWAVE PLASMA TORCH ATOMIC-EMISSION SPECTROMETRY, Spectrochimica acta, Part B: Atomic spectroscopy, 53(10), 1998, pp. 1427-1435
A thermospray nebulizer was used as a sample introduction device for m
icrowave plasma torch (MPT) atomic emission spectrometry (AES). Experi
mental parameters, including the power supplied to the MPT, the flow r
ates of support and carrier gases, the observation height, the sample
uptake rate, the thermospray working temperature, the temperature of t
he aerosol spray chamber and cooling water were optimized. Under optim
um conditions, the relative standard deviation (RSD) of 10 measurement
s for 21 elements is in the range 0.3-2.0%. The detection limits were
improved in comparison with the ultrasonic nebulizer as sample introdu
ction technique for MPT-AES. The inclusion of 20% methanol into the MP
T showed there is no effect on the stability of MPT discharge. The tec
hnique can thus be held to have the potential for interface to reverse
-phase HPLC systems. (C) 1998 Elsevier Science B.V. All rights reserve
d.