P. Westbroek et al., Measurement and control of hydrogen peroxide concentration in alkaline solution by means of amperometric sensor system, ANALYT CHIM, 385(1-3), 1999, pp. 423-428
An amperometric sensor system has been developed to measure and control the
hydrogen peroxide concentration in a continuous and in-line way, over a wi
de concentration range (0.1-70 g/l), with high precision, accuracy and sele
ctivity and over a long period of time without the need of recalibration or
replacement of the active part of the sensor system. The system performs w
ell in an industrial environment under the condition that the pH of the pro
cess liquid is at least 10.5. (C) 1999 Elsevier Science B.V. All rights res
erved.