A flow-through amperometric immunofiltration assay system based on disposab
le porous filter-membranes for rapid detection of Escherichia coli O157:H7
has been developed. The analytical system utilizes flow-through, immunofilt
ration and enzyme immunoassay techniques in conjunction with an amperometri
c sensor. The parameters affecting the immunoassay such as selection of app
ropriate filter membranes, membrane pore size, antibody binding capacity an
d the concentrations of immunoreagents were investigated and optimized. Non
-specific adsorption of the enzyme conjugate was investigated and minimized
. A sandwich scheme of immunoassay was employed and the immunofiltration sy
stem allows to specifically and directly detect E. coli cells with a lower
detection limit of 100 cells/ml. The working range is from 100 to 600 cells
/ml with an overall analysis time of 30 min. No pre-enrichment was needed.
This immunosensor can be easily adapted for assay of other microorganisms a
nd may be a basis for a new class of highly sensitive bioanalytical devices
for rapid quantitative detection of bacteria. (C) 1999 Elsevier Science S.
A. All rights reserved.