Background-Endocardial mapping of sustained arrhythmias has traditionally b
een performed with a roving diagnostic catheter. Although this approach is
adequate for many tachyarrhythmias, it has limitations, The purpose of-this
study was to evaluate a novel noncontact mapping system for assessing atri
al tachyarrhythmias.
Methods and Results-The mapping system consists of a 9F multielectrode-arra
y balloon catheter that has 64 active electrodes and ring electrodes for em
itting a locator signal. The locator signal was used to construct a 3-dimen
sional right atrial map; it was independently validated and was highly accu
rate. Virtual electrograms were calculated at 3360 endocardial sites in the
right atrium. We evaluated right atrial activation by positioning the ball
oon catheter in the mid right atrium via a femoral venous approach. Experim
ents were performed on 12 normal mongrel dogs. The mean correlation coeffic
ient between contact and virtual electrograms was 0.80+/-0.12 during sinus
rhythm. Fifty episodes of atrial flutter induced in Il animals were evaluat
ed. In the majority of experiments, complete or almost complete reentrant c
ircuits could be identified within the right atrium. Mean correlation coeff
icient between virtual and contact electrograms was 0.85+/-0.17 in atrial f
lutter. One hundred fifty-six episodes of pacing-induced atrial fibrillatio
n were evaluated in 11 animals. Several distinct patterns of right atrial a
ctivation were seen, including single-activation wave fronts and multiple s
imultaneous-activation wave fronts. Mean correlation coefficient between vi
rtual and contact electrograms during atrial fibrillation was 0.81+/-0.18.
The accuracy of electrogram reconstruction was lower at sites >4.0 cm from
the balloon center and at sites with a high spatial complexity of electrica
l activation.
Conclusions-This novel noncontact mapping system can evaluate conduction pa
tterns during sinus rhythm, demonstrate reentry during atrial flutter, and
describe right atrial activation during atrial fibrillation. The accuracy o
f electrogram reconstruction was good at sites <4.0 cm from the balloon cen
ter, and thus the system has the ability to perform high-resolution multisi
te mapping of atrial tachyarrhythmias in vivo.