T-1 weighted sequences for hepatic MRI: Re-evaluation using a phased arraycoil

Citation
Dr. Decorato et al., T-1 weighted sequences for hepatic MRI: Re-evaluation using a phased arraycoil, CLIN IMAG, 23(1), 1999, pp. 26-31
Citations number
13
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging
Journal title
CLINICAL IMAGING
ISSN journal
08997071 → ACNP
Volume
23
Issue
1
Year of publication
1999
Pages
26 - 31
Database
ISI
SICI code
0899-7071(199901/02)23:1<26:TWSFHM>2.0.ZU;2-T
Abstract
The purpose of this study was to compare three T1-weighted sequences for he patic magnetic resonance (MRI) imaging with the use of a body phased array coil. Three different T1-weighted MR sequences were compared: a conventiona l spin echo (CSE); half-Fourier spin echo (HFSE), and a gradient recalled e cho (GRE). Three independent reviewers compared the sequences both quantita tively and qualitatively, The T1-weighted GRE sequence scored highest for o verall image quality (p < 0.001), lesion conspicuity (p = 0.012), and yield ed the highest contrast to noise (C/N) values. GRE T1-weighted images are t he best for hepatic MRI. (C) Elsevier Science Inc., 1999.