Cj. Bates, Upstream installation and misalignment effects on the performance of a modified electromagnetic flowmeter, FLOW MEAS I, 10(2), 1999, pp. 79-89
The performance of a modified Danfoss 50 mm diameter electromagnetic flowme
ter has been investigated when installed downstream from three different pi
pe diameters-50 mm, 55 mm and 45 mm. The effects of a 3 mm misalignment of
the flowmeter, in both the vertical and horizontal planes, with respect to
each of the three upstream pipe diameters has also been identified. The lar
gest percentage errors are reported for the 45 mm upstream diameter pipe, w
ith the flowmeter misaligned by 3 mm in the horizontal plane. The vertical
and horizontal mean velocity and root-mean-square velocity profiles, measur
ed within the flowmeter using laser Doppler anemometry, show significant va
riations in comparison with the ideal, fully developed profiles. (C) 1999 E
lsevier Science Ltd. All rights reserved.