T. Kojima et al., Uncertainty estimation in Co-60 gamma-ray dosimetry at JAERI involving a two-way dose intercomparison study with NPL in the dose range 1-50 kGy, RADIAT PH C, 54(6), 1999, pp. 619-626
The uncertainties in Co-60 gamma-ray dosimetry at JAERI have been estimated
by reviewing irradiation and calibration procedures. The uncertainty in do
se delivered by the JAERI calibration system has been estimated to be 2.2%
(2 sigma) and the uncertainty in dose measurement using JAERI alanine/ESR t
ransfer dosimeters has been estimated to be 3.4% (2 sigma), A two-way dose
intercomparison study between NPL and JAERI employing alanine-PS, alanine-p
araffin, and dichromate dosimeters as transfer dosimeters was performed to
confirm the estimated uncertainty over the dose range 1-50 kGy. Dose interc
omparison results using both alanine and dichromate dosimeters show agreeme
nt within about 2%, which gives confidence in the estimated uncertainty in
Co-60 gamma-ray dosimetry at JAERI. (C) 1999 Elsevier Science Ltd. All righ
ts reserved.