U. Merker et al., High-resolution CO-laser sideband spectrometer for molecular-beam optothermal spectroscopy in the 5-6.6 mu m wavelength region, REV SCI INS, 70(4), 1999, pp. 1933-1938
We have set up a new CO-laser sideband spectrometer for high-resolution mol
ecular-beam optothermal spectroscopy in the mid infrared. By mixing CO-lase
r lines with tunable microwave radiation, a spectral coverage of about 50%
is achieved. Using a microwave resonator, a typical output power of 3 mW is
realized in the region of 5-6.6 mu m (1500-2000 cm(-1)). The resolution in
our molecular-beam apparatus with optothermal detection is 2.8 MHz (full w
idth at half maximum). The new setup allowed the observation of the first h
igh-resolution rotationally resolved spectrum of formic acid dimer (HCOOH)(
2). (C) 1999 American Institute of Physics. [S0034-6748(99)02204-2].