Polycapillary x-ray optic spectral gain and transmission

Citation
J. Gormley et al., Polycapillary x-ray optic spectral gain and transmission, X-RAY SPECT, 28(2), 1999, pp. 115-120
Citations number
19
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
X-RAY SPECTROMETRY
ISSN journal
00498246 → ACNP
Volume
28
Issue
2
Year of publication
1999
Pages
115 - 120
Database
ISI
SICI code
0049-8246(199903/04)28:2<115:PXOSGA>2.0.ZU;2-F
Abstract
Polycapillary x-ray optics are gaining increased attention as their transmi ssion characteristics and manufacturing difficulties are better understood. This paper reports experimental results obtained with an optic that was de signed for use with both Mo K alpha (17.5 keV) and Cu K alpha (8 keV), The spectral gain function from 4 to 30 keV was determined from measurements ma de with an Si(Li) detector, and the results were compared with Monte Carlo simulations. The spatial variation in the spectral transmission was examine d with both an Si(Li) detector and a magnifying x-ray CCD camera. The resul ts show that even though the individual capillaries are tapered, their perf ormance is well described by a simple parabolic relationship between the ra dius of curvature and the maximum energy transmitted derived for capillarie s of constant diameter. Copyright (C) 1999 John Wiley & Sons, Ltd.