B. Meyns et al., MINIATURIZED IMPLANTABLE ROTARY BLOOD PUMP IN ATRIAL-AORTIC POSITION SUPPORTS AND UNLOADS THE FAILING HEART, Cardiovascular surgery, 6(3), 1998, pp. 288-295
The purpose of this study was to test the pump performance of a miniat
urized, newly developed, implantable rotary blood pump (the diagonal p
ump). In a series of six sheep, the pump was interposed in an atrial-a
ortic connection. Heart failure was induced by serial injection of gla
ss beads into the left coronary artery, The assistance of the non-fail
ing heart (with pump output up to 4.5 liter/min) did not change the to
tal cardiac output or the blood pressure of the animal. The heart was
significantly unloaded as demonstrated by a drop in first derivative o
f the left ventricular pressure (dP/dt) max (from 1645 to 1113 mmHg/s;
P = 0.0003). Because of the specific interaction between heart functi
on and rotary blood pump performance, the pump provided considerably m
ore flow in the failing than in the non-failing heart. Cardiac output
and perfusion pressures were restored to pre-assist values in the fail
ing heart model. The heart-rotary blood pump interaction is an ideal s
ubstrate for long-term assistance. With this miniaturized rotary blood
pump, hemodynamics in a severely failing heart can be restored. Atria
l cannulation, which leaves the left ventricle untouched, provides sim
ilar results as apical cannulation in the failing heart. (C) 1998 The
International Society for Cardiovascular Surgery. Published by Elsevie
r Science Ltd. All rights reserved.