SIMPLE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR DETERMINATION OF KETOCONAZOLE IN HUMAN PLASMA

Authors
Citation
Kh. Yuen et Kk. Peh, SIMPLE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR DETERMINATION OF KETOCONAZOLE IN HUMAN PLASMA, Journal of chromatography B. Biomedical sciences and applications, 715(2), 1998, pp. 436-440
Citations number
13
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Journal title
Journal of chromatography B. Biomedical sciences and applications
ISSN journal
13872273 → ACNP
Volume
715
Issue
2
Year of publication
1998
Pages
436 - 440
Database
ISI
SICI code
0378-4347(1998)715:2<436:SHLMFD>2.0.ZU;2-R
Abstract
A simple high-performance liquid chromatographic method using fluoresc ence detection was developed for the determination of ketoconazole in human plasma. The method entailed direct injection of the plasma sampl e after deproteinization using acetonitrile. The mobile phase comprise d 0.05 M disodium hydrogen orthophosphate and acetonitrile (50:50, v/v ) adjusted to pH 6. Analysis was run at a flow-rate of 1.5 ml/min with the detector operating at an excitation wavelength of 260 nm and an e mission wavelength of 375 nm. The method is specific and sensitive wit h a quantification limit of approximately 60 ng/ml and a detection Lim it of 40 ng/ml at a signal-to-noise ratio of 3:1. Mean absolute recove ry value was about 105%, while the within-day and between-day coeffici ent of variation and percent error values of the assay method were all less than 14%. The calibration curve was linear over a concentration range of 62.5-8000 ng/ml. (C) 1998 Elsevier Science B.V. All rights re served.