RELIABILITY OF AN AUTOMATIC SENSING TEST WITH BEAT-TO-BEAT DISPLAY OFTHE SIGNAL AMPLITUDE

Citation
A. Lazarus et al., RELIABILITY OF AN AUTOMATIC SENSING TEST WITH BEAT-TO-BEAT DISPLAY OFTHE SIGNAL AMPLITUDE, PACE, 21(10), 1998, pp. 1881-1884
Citations number
3
Categorie Soggetti
Cardiac & Cardiovascular System","Engineering, Biomedical
Journal title
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
ISSN journal
01478389 → ACNP
Volume
21
Issue
10
Year of publication
1998
Pages
1881 - 1884
Database
ISI
SICI code
0147-8389(1998)21:10<1881:ROAAST>2.0.ZU;2-N
Abstract
In contrast to testing of the capture threshold, the reliability of se nsing tests has been little studied. This study was performed to test the automatic sensing algorithm included in the Biotronik pacing syste ms. The automatic measurements made by the devices were compared with the direct manual measurements made of 271 atrial (72%) or ventricular (28%) electrograms recorded in ten patients. A high correlation (r = 0.995, P < 0.0001) was found between the two types of measurements. Th e use of this function should facilitate the verification of accurate sensing during the long-term follow-up of pacemaker patients, and offe rs a tool to analyze the variations in amplitude of intracardiac signa ls.