Objective-To evaluate maintenance of proper VDD function, defined as p
ersistence of sinus rhythm with atrial synchronous ventricular pacing,
and to define factors predicting failure of the VDD mode in patients
with atrioventricular (AV) block and normal sinus function. Design-Obs
ervational study in 86 consecutive patients (mean (SD) age 74 (12) yea
rs; 38 women, 48 men) with single lead VDD pacing systems (Intermedics
Unity, n = 66, Medtronic Thera VDD, n = 20), implanted for high degre
e AV block with documented normal sinus node. Pacemaker function was a
ssessed by event counters, telemetric measurements, and Holter recordi
ngs. Demographic, radiological, and pacing variables were correlated w
ith loss of proper VDD function. Results-During a mean (SD) follow up
of 10 (10) months (range 1-37), sinus rhythm and atrial triggered vent
ricular pacing were maintained in 70 of 86 patients (81%). Atrial unde
rsensing was observed in nine patients, lead migration in two, atrial
fibrillation in three, and symptomatic sinus bradycardia in two. Univa
riate predictors of loss of groper VDD function were: low position of
the atrial dipole relative to the carina (greater than or equal to 6 c
m; p < 0.01) during fluoroscopy; and maximum programmable atrial sensi
tivity of the pacemaker (p = 0.03). Hn a multivariate analysis, only d
ipole position remained predictive of outcome (p < 0.02). Not predicti
ve: were sex, ape, symptoms before pacemaker implantation, cardiothora
cic ratio or dilation of individual heart chambers on chest x ray, sid
e of device implant, and P wave amplitude at implant. Conclusions-To m
aintain proper VDD function ifs the long term, a low anatomical dipole
position relative to the carina should be avoided. Electrical. guidan
ce of dipole positioning does not seem to influence long term outcome.