T. Akizawa et al., DEVELOPMENT OF A NEW CELLULOSE TRIACETATE MEMBRANE WITH A MICROGRADIENT POROUS STRUCTURE FOR HEMODIALYSIS, ASAIO journal, 44(5), 1998, pp. 584-586
Dialytic performance and biocompatibility of a newly developed cellulo
se triacetate (CTA) membrane with a microgradient porous structure, pr
oduced without using polyvinylpyrrolidone (PVP) and liquid paraffin, w
ere compared with those of a conventional polysulfone (PS) membrane di
alyzer. In vitro and clinical results demonstrated no significant diff
erence in the dialytic performance and biocompatibility of the two dia
lyzers, but the CTA dialyzer lost less albumin during dialysis than th
e PS. These results suggest that a CTA membrane dialyzer with a porous
microgradient structure attained comparable performance and biocompat
ibility to PS, and the risk of albumin loss was suppressed by the new
CTA membrane.