L. Poller et al., THE EUROPEAN CONCERTED ACTION ON ANTICOAGULATION (ECAA) - FIELD STUDIES OF COAGULOMETER EFFECTS ON THE ISI OF ECAA THROMBOPLASTINS, Thrombosis and haemostasis, 80(4), 1998, pp. 615-623
Local calibration studies have been performed with lyophilized plasmas
at 155 European laboratories to assess coagulometer effects on manual
ISI of ECAA thromboplastins (low ISI human recombinant and high ISI r
abbit brain). Common sets of 7 normal and 20 artificially depleted lyo
philized plasmas were tested with the thromboplastins in the routine l
ocal coagulometers. With the human reagent, marked lowering of the man
ual ISI resulted with most coagulometers, which was associated with di
sproportionate shortening of the normals. Where this shortening did no
t occur, there was little coagulometer effect on the ISI. With the rab
bit reagent, proportionate shortening with both normal and abnormal pl
asmas occurred in most instruments with little effect on ISI. INR corr
ection by local ISI assignment appeared successful with the ECAA human
reagent. There was negligible INR correction from local calibration o
f the ECAA rabbit reagent in coagulometers where thromboplastin ISI we
re unchanged from the ECAA established manual values.