Le. Flores et al., AUTOMATIC SYSTEM FOR THE DIRECT AND CONTINUOUS MEASUREMENT OF THE IRREVERSIBILITY LINE OF HIGH T-C SUPERCONDUCTORS, Review of scientific instruments, 69(10), 1998, pp. 3634-3636
A system for the direct and continuous determination of the irreversib
ility line in high T-c superconductors is described. The system, based
on a standard ac susceptometer, operates by applying a controlled dc
field to the sample that maximizes the imaginary (out-of-phase) compon
ent of the ac susceptibility. This control is achieved using a proport
ional-integral-derivative loop with the reference set to zero and with
the magnetic field derivative of the imaginary component of the ac su
sceptibility taken as the feedback sample (error signal). To obtain a
dc voltage proportional to this derivative a double lock-in detection
scheme is used. This apparatus is able to carry out a continuous measu
rement of this line by sweeping the temperature in the range of intere
st while the sample is maintained, by the controller, at the irreversi
bility magnetic field at each temperature. Hence, the whole process is
accomplished in the same time as a typical resistivity-versus-tempera
ture measurement, which is faster than other methods. (C) 1998 America
n Institute of Physics. [S0034-6748(98)02408-3].