IN-PHANTOM NEUTRON FLUENCE MEASUREMENTS IN THE ORTHOGONAL BIRMINGHAM BORON NEUTRON-CAPTURE THERAPY BEAM

Citation
Da. Tattam et al., IN-PHANTOM NEUTRON FLUENCE MEASUREMENTS IN THE ORTHOGONAL BIRMINGHAM BORON NEUTRON-CAPTURE THERAPY BEAM, Medical physics, 25(10), 1998, pp. 1964-1966
Citations number
6
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
00942405
Volume
25
Issue
10
Year of publication
1998
Pages
1964 - 1966
Database
ISI
SICI code
0094-2405(1998)25:10<1964:INFMIT>2.0.ZU;2-H
Abstract
This paper presents the results of an experimental investigation into the performance of the Birmingham accelerator-based epithermal BNCT be am. In-phantom gold foil activation and boron trifluoride tube measure ments have been used. The results have been compared with calculated r esponse rates using Monte Carlo modeling of the entire neutron system from source to phantom and detector. The excellent agreement obtained gives us confidence in the validity of the simulations and our ability to predict accurately the neutronic performance of our BNCT facility. (C) 1998 American Association of Physicists in Medicine. [S0094-2405( 98)00210-7].