HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF OXALIC-ACID IN TEA USING TRIS(1,10-PHENANTHROLINE)-RUTHENIUM(II) CHEMILUMINESCENCE

Citation
Fw. Wu et al., HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF OXALIC-ACID IN TEA USING TRIS(1,10-PHENANTHROLINE)-RUTHENIUM(II) CHEMILUMINESCENCE, Analytical sciences, 14(5), 1998, pp. 971-973
Citations number
13
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
09106340
Volume
14
Issue
5
Year of publication
1998
Pages
971 - 973
Database
ISI
SICI code
0910-6340(1998)14:5<971:HLDOOI>2.0.ZU;2-R
Abstract
The chemiluminescence (CL) of Ru(phen)(3)(2+) was applied to HPLC dete rmination of oxalic acid, which was separated by a C-18 reverse-phase column with a mobile phase of 0.02 mol/l NH4Ac, mixed with 0.25 mol/l Ru(phen)(3)(2+) and 2.0 mmol/l Ce(SO4)(2) (0.08 mol/l H2SO4), and then passed through a homemade CL detector. The reaction of Ce(IV)oxidized Ru(phen)(3)(2+) and the oxalic acid emitted light. The detection limi t was 6.2x10(-6) mol/l for oxalic acid at a S/N ratio of 3, the relati ve standard deviation for 5 replicate injections of 1x10(-3) mol/l oxa lic acid standard was calculated as 5.6%, and the linear calibration r ange was 1x10(-5) to 4x10(-3) mol/l. The method was successfully appli ed to determination of oxalic acid in tea.