G. Stevens et al., THYROID DOSE IN CHILDREN UNDERGOING PROPHYLACTIC CRANIAL IRRADIATION, International journal of radiation oncology, biology, physics, 42(2), 1998, pp. 385-390
Citations number
19
Categorie Soggetti
Oncology,"Radiology,Nuclear Medicine & Medical Imaging
Purpose: To determine the radiation dose received by the thyroid gland
as a result of prophylactic cranial irradiation (PCI) in childhood le
ukemia and the factors influencing that dose. Methods and Materials: T
he dose to the thyroid resulting from simulated cranial irradiation wi
th parallel opposed lateral fields of an adult anthropomorphic (ART) p
hantom with both 6 MV X-rays and Cobalt-60 gamma-rays was measured usi
ng thermoluminescent dosimeters (TLDs). The dependence of thyroid dose
on the distance of the field from the thyroid and the proportions of
thyroid dose from stray radiation (leakage, scatter from jaws, etc.) a
nd tissue scattered radiation were measured. The effects of a shadow t
ray and shielding blocks mere also determined. Calculation of thyroid
dose using the Clarkson scatter integration method was performed for 6
MV X-rays to compare with the measured doses. In vivo thyroid dose es
timates were made using TLD measurements for three children receiving
PCI with 6 MV X-rays. Results: Using open, unshielded fields, the thyr
oid region of the phantom received 1.2-1.4% of the prescribed cranial
dose for 6 MV X-rays and 1.5-1.7% for Cobalt-60. For both treatment un
its, stray radiation accounted for approximately two thirds of the thy
roid dose and tissue scatter accounted for the remaining one third. Th
e thyroid dose increased as the field moved closer to the thyroid, wit
h an increasing proportion of the dose due to tissue scatter. Placemen
t of a thyroid shielding block on a shadow tray reduced the thyroid do
se by only 20% compared with the open, unshielded setup. Thyroid dose
from 6 MV using open fields was affected by the orientation of the col
limator. When the inferior field edge was defined by the lower jaw, th
e dose was reduced by 27% compared with the upper jaw. Good correlatio
n of dose to the thyroid region was obtained between phantom measured
doses, in vivo measured doses and calculation of dose using the Clarks
on method. Conclusion: For PCI doses of 1800 or 2400 cGy in the adult
phantom, the dose to the thyroid was 20-40 cGy (1-2%). For small child
ren this could rise to approximately 5% of the prescribed dose, of whi
ch half was due to stray radiation. As the thyroid in children is very
sensitive to radiation and the dose-response curve for thyroid tumor
induction is linear, attempts to shield the thyroid during cranial irr
adiation are mandatory. Cobalt-60 units should not be used, as the thy
roid dose was higher than using 6 MV X-rays. Collimator orientation an
d the use of shadow trays and shielding were important factors in dete
rmining thyroid dose. (C) 1998 Elsevier Science Inc.