COMPARISON OF LUMPED-ELEMENT AND TRANSMISSION-LINE MODELS FOR THICKNESS-SHEAR-MODE QUARTZ RESONATOR SENSORS

Citation
Rw. Cernosek et al., COMPARISON OF LUMPED-ELEMENT AND TRANSMISSION-LINE MODELS FOR THICKNESS-SHEAR-MODE QUARTZ RESONATOR SENSORS, IEEE transactions on ultrasonics, ferroelectrics, and frequency control, 45(5), 1998, pp. 1399-1407
Citations number
10
Categorie Soggetti
Engineering, Eletrical & Electronic",Acoustics
ISSN journal
08853010
Volume
45
Issue
5
Year of publication
1998
Pages
1399 - 1407
Database
ISI
SICI code
0885-3010(1998)45:5<1399:COLATM>2.0.ZU;2-2
Abstract
Both a transmission-line model and its simpler variant, a lumped-eleme nt model, can be used to predict the responses of a thickness-shear-mo de quartz resonator sensor. Relative deviations in the parameters comp uted by the two models (shifts in resonant frequency and motional resi stance) do not exceed 3% for most practical sensor configurations oper ating at the fundamental resonance. If the ratio of the load surface m echanical impedance to the quartz shear characteristic impedance does not exceed 0.1, the lumped-element model always predicts responses wit hin 1% of those for the transmission-line model.