T. Usuda et al., THE METHODS FOR THE CALIBRATION OF VIBRATION PICKUPS BY LASER INTERFEROMETRY - PART III - PHASE-LAG EVALUATION, Measurement science & technology (Print), 9(10), 1998, pp. 1672-1677
The calibration of vibration pick-ups by the sine-approximation method
has been examined. This method which uses a modified Michelson-type l
aser interferometer has been proposed as an ISO draft standard. The ev
aluation procedure, the experimental set-up and calculation of the unc
ertainty in the phase lag are reported. A piezoelectric-type vibration
pick-up has been evaluated experimentally. Expanded uncertainties of
less than +/-5 x 10(-3) rad in phase-lag evaluation have been obtained
in the experiment.