SIMULTANEOUS DETERMINATION OF FLAVONOIDS, PHENOLIC-ACIDS, AND COUMARINS IN 7 MEDICINAL SPECIES BY HPLC DIODE-ARRAY DETECTOR/

Citation
Pb. Andrade et al., SIMULTANEOUS DETERMINATION OF FLAVONOIDS, PHENOLIC-ACIDS, AND COUMARINS IN 7 MEDICINAL SPECIES BY HPLC DIODE-ARRAY DETECTOR/, Journal of liquid chromatography & related technologies, 21(18), 1998, pp. 2813-2820
Citations number
7
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
ISSN journal
10826076
Volume
21
Issue
18
Year of publication
1998
Pages
2813 - 2820
Database
ISI
SICI code
1082-6076(1998)21:18<2813:SDOFPA>2.0.ZU;2-W
Abstract
A simple and accurate reversed phase HPLC procedure is proposed for th e determination of 19 phenolic compounds (flavonoids, phenolic acids, and coumarins) in seven medicinal species. The sample preparation invo lved extraction, alkaline and acid hydrolysis, and purification throug h a Bond-Elut C18 column. The chromatographic separation was achieved using a reversed phase column Spherisorb ODS2 (5 mu m; 25.0 x 0.46 cm) . Gradient elution was carried out using water-formic acid (19:1) (A) and methanol (B). A diode-array detector monitored the effluent and ch romatograms were recorded at 280, 320, and 350 nm. The optimised metho dology seems to be useful for the phytochemical analysis of plant extr acts. A close correlation between the phenolic compound patterns and t he botanical origin of plants was found.