CALCULATION OF DEPTH-DOSE AND DOSE PER MONITOR UNIT FOR IRREGULARLY SHAPED ELECTRON FIELDS

Citation
Fm. Khan et al., CALCULATION OF DEPTH-DOSE AND DOSE PER MONITOR UNIT FOR IRREGULARLY SHAPED ELECTRON FIELDS, Physics in medicine and biology (Print), 43(10), 1998, pp. 2741-2754
Citations number
18
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
00319155
Volume
43
Issue
10
Year of publication
1998
Pages
2741 - 2754
Database
ISI
SICI code
0031-9155(1998)43:10<2741:CODADP>2.0.ZU;2-O
Abstract
A new dosimetric quantity, the lateral build-up ratio (LBR), has been introduced to calculate depth dose distribution for any shaped field. Factors to account for change in incident fluence with collimation are applied separately. The LBR data for a small circular field are used to extract radial spread of the pencil beam, sigma(r), as a function o f depth and energy. By using the relationship between LBR, sigma(r), e nergy and depth, a formalism is developed to calculate dose per monito r unit for any shaped field. Criteria for lateral scatter equilibrium are also developed which are useful in clinical dosimetry.