Se. Urbak et al., ULTRASOUND BIOMICROSCOPY - II - INTRAOBSERVER AND INTEROBSERVER REPRODUCIBILITY OF MEASUREMENTS, Acta ophthalmologica Scandinavica, 76(5), 1998, pp. 546-549
Purpose: To evaluate intraobserver and interobserver reproducibility o
f measurements in images obtained by the 50 MHz Humphrey(R) Ultrasound
Biomicroscope. Methods: The first stored image of central cornea, cen
tral anterior chamber, and angle structures obtained at 50 examination
s were measured twice by three observers. Nine different parameters me
re measured in angle images and two in images of central cornea and an
terior chamber. Intraobserver reproducibility was assessed by calculat
ing the coefficient of variation and interobserver reproducibility by
a two-ways ANOVA. Results: Intraobserver reproducibility was high for
all measurements of central cornea thickness and anterior chamber dept
h with a coefficient of variation (%) less than or equal to 3.8%. In a
ngle images the intraobserver reproducibility of measurements,vas ofte
n lower (coefficient of variation (%) 1.3-52.6%). The lower reproducib
ility was most noticeable in measurements involving the less defined s
tructures. Interobserver reproducibility was poor. Conclusions: Compar
ison of measurements are best done of well defined structures and by o
nly one observer.