Purpose: Determination of the accurate interval between the measuremen
ts in repeated tonometry, Methods: Two hundred-and-seventy subjects we
re randomly divided into 6 groups. Two successive intraocular pressure
(IOP) measurements were performed with Goldmann-type applanation tono
meter at different time intervals in each group. The time intervals be
tween the two measurements were 1, 2, 3, 4, 5 and 10 minutes. The resu
lts in each group were statistically analyzed with the significance te
st for matched observations, Results: We recorded a statistically sign
ificant difference between the first and the second IOP measurements a
t 1, 3, 4 and 5-minute-intervals. The differences were not significant
at 2 and 10 minute-groups. Conclusion: The time interval between succ
essive IOP measurements should be 2 or 10 minutes for accurate tonomet
ry.