Av. Pereira et O. Fatibello, SPECTROPHOTOMETRIC FLOW INJECTION DETERMINATION OF L-ASCORBIC-ACID WITH A PACKED REACTOR CONTAINING FERRIC HYDROXIDE, Talanta, 47(1), 1998, pp. 11-18
A flow injection system with spectrophotometric detection is proposed
for determining L-ascorbic acid in pharmaceutical formulations. In thi
s system a column containing Fe(OH)(3) immobilized in polyester resin
(packed reactor) is inserted before the detector. Fe(III)-1,10-phenant
hroline complex is reduced by L-ascorbic acid to produce Fe(II)-1,10-p
henanthroline complex which is monitored at 510 nn. Under the optimum
analytical conditions, the linearity of the calibration equation for L
-ascorbic acid ranged from 5.0 x 10(-6) to 6.0 x 10(-5) M of added amo
unt. The detection limit was 5.0 x 10(-7) M and recoveries between 98.
5-102.0% were obtained. No interference was observed from the common e
xcipients of pharmaceutical formulations and other active substances s
uch as acetylsalicylic acid, caffeine and thiamine. (C) 1998 Elsevier
Science B.V. All rights reserved.