EVALUATION OF THE PRECISION AND ACCURACY OF AN AUTOMATED SAMPLE INTRODUCTION ACCESSORY FOR THE FLAME ATOMIC-ABSORPTION SPECTROMETRIC MEASUREMENT OF CALCIUM IN SERUM
Ms. Epstein et al., EVALUATION OF THE PRECISION AND ACCURACY OF AN AUTOMATED SAMPLE INTRODUCTION ACCESSORY FOR THE FLAME ATOMIC-ABSORPTION SPECTROMETRIC MEASUREMENT OF CALCIUM IN SERUM, Talanta, 47(1), 1998, pp. 95-102
The measurement precision and accuracy that results from the use of a
sample introduction peristaltic pump system for automated instrument c
alibration and analysis is compared to manual methods based on traditi
onal gravimetric and volumetric dilution procedures for flame atomic a
bsorption analyses in water and serum matrices. Whilst use of the auto
mated system improves speed of analysis, measurement precision was fou
nd to be approximately 2-fold worse than the manual methods based on g
ravimetric and volumetric dilution procedures. (C) 1998 Elsevier Scien
ce B.V. All rights reserved.