EVALUATION OF THE PRECISION AND ACCURACY OF AN AUTOMATED SAMPLE INTRODUCTION ACCESSORY FOR THE FLAME ATOMIC-ABSORPTION SPECTROMETRIC MEASUREMENT OF CALCIUM IN SERUM

Citation
Ms. Epstein et al., EVALUATION OF THE PRECISION AND ACCURACY OF AN AUTOMATED SAMPLE INTRODUCTION ACCESSORY FOR THE FLAME ATOMIC-ABSORPTION SPECTROMETRIC MEASUREMENT OF CALCIUM IN SERUM, Talanta, 47(1), 1998, pp. 95-102
Citations number
8
Categorie Soggetti
Chemistry Analytical
Journal title
Talanta
ISSN journal
00399140 → ACNP
Volume
47
Issue
1
Year of publication
1998
Pages
95 - 102
Database
ISI
SICI code
0039-9140(1998)47:1<95:EOTPAA>2.0.ZU;2-P
Abstract
The measurement precision and accuracy that results from the use of a sample introduction peristaltic pump system for automated instrument c alibration and analysis is compared to manual methods based on traditi onal gravimetric and volumetric dilution procedures for flame atomic a bsorption analyses in water and serum matrices. Whilst use of the auto mated system improves speed of analysis, measurement precision was fou nd to be approximately 2-fold worse than the manual methods based on g ravimetric and volumetric dilution procedures. (C) 1998 Elsevier Scien ce B.V. All rights reserved.