ROLE OF CATALYST TO IMPROVE P-10 FILLED GAS COUNTER PERFORMANCE

Citation
B. Behera et al., ROLE OF CATALYST TO IMPROVE P-10 FILLED GAS COUNTER PERFORMANCE, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 416(1), 1998, pp. 81-84
Citations number
3
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
416
Issue
1
Year of publication
1998
Pages
81 - 84
Database
ISI
SICI code
0168-9002(1998)416:1<81:ROCTIP>2.0.ZU;2-W
Abstract
wThe long-term performance of a Delta E gas counter filled with P-10 g as at a pressure of 130 Torr, and operated in the proportional mode, i s monitored. The presence of iron-oxide based catalysts within the cou nter volume successfully removes electronegative impurities from outga ssing and initial contaminants. It has been possible to operate this c ounter continuously for more than three months, with good resolution ( 9-10%). The initial slow decrease of peak channel with time has been a ttributed to the dynamic increase of electronegative impurities. It ha s been concluded that the long-term performance of a gas counter depen ds substantially on the concentration of the catalysts. (C) 1998 Elsev ier Science B.V. All rights reserved.