CYCLODEXTRIN-BASED OPTOSENSOR FOR THE DETERMINATION OF QUININE

Authors
Citation
Zl. Gong et Zj. Zhang, CYCLODEXTRIN-BASED OPTOSENSOR FOR THE DETERMINATION OF QUININE, Fresenius' journal of analytical chemistry, 357(8), 1997, pp. 1093-1096
Citations number
19
Categorie Soggetti
Chemistry Analytical
ISSN journal
09370633
Volume
357
Issue
8
Year of publication
1997
Pages
1093 - 1096
Database
ISI
SICI code
0937-0633(1997)357:8<1093:COFTDO>2.0.ZU;2-F
Abstract
A flow-through optosensor for quinine is described. It has been develo ped in conjunction with a flow-injection analysis system and uses immo bilized beta-cyclodextrin as the sensing agent. The detection limit fo r quinine was 0.20 ng ml(-1) with a relative standard deviation of 2.1 % for the determination of 0.05 mu g ml(-1) (n = 7) of quinine. The me thod has been successfully tested for the determination of quinine in some pharmaceutical preparations.