Background-The American Thoracic Society (ATS) has set the acceptable
resistance far spirometers at less than 1.5 cm H2O/1/s over the flow r
ange 0-14 1/s and for monitoring devices at less than 2.5 cm H2O/1/s (
0-14 1/s). The aims of this study were to determine the resistance cha
racteristics of commonly used spirometers and monitoring devices and t
he effect of resistance on ventilatory function. Methods-The resistanc
e of five spirometers (Vitalograph wedge bellows, Morgan rolling seal,
Stead Wells water sealed, Fleisch pneumotachograph, Lilly pneumotacho
graph) and three monitoring devices (Spiro 1, Ferraris, mini-Wright) w
as measured from the back pressure developed over a range of known flo
ws (1.6-13.1 1/s). Peak expiratory flow. (PEF), forced expiratory flow
in one second (FEV1), forced vital capacity (FVC), and mid forced exp
iratory flow (FEF25-75%) were measured on six subjects with normal lun
g function and 13 subjects with respiratory disorders using a pneumota
chograph. Ventilatory function was then repeated with four different s
ized resistors (approximately 1-11 cmH(2)O/1/s) inserted between the m
outhpiece and pneumotachograph. Results-All five diagnostic spirometer
s and two of the three monitoring devices passed the ATS upper limit f
or resistance. PEF, FEV1 and FVC showed significant (p < 0.05) inverse
correlations with added resistance with no significant difference bet
ween the normal and patient groups. At a resistance of 1.5 cm H2O/1/s
the mean percentage falls (95% confidence interval) were: PEF 6.9% (5.
4 to 8.3); FEV1 1.9% (1.0 to 2.8), and FVC 1.5% (0.8 to 2.3). Conclusi
ons-The ATS resistance specification for diagnostic spirometers appear
s to be appropriate. However, the specification for monitoring devices
may be too conservative. PEF was found to be the most sensitive index
to added resistance.