Kl. Giboni et E. Aprile, EVALUATION OF CDTE DETECTORS WITH SCHOTTKY CONTACTS FOR IMAGING APPLICATIONS, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 416(2-3), 1998, pp. 319-327
A CdTe detector with Schottky contacts has been tested with special em
phasis on the requirements for imaging arrays. The detector can withst
and a much higher field strength than conventional CdTe detectors with
two ohmic contacts. The higher field strength results in complete cha
rge collection, including the holes which are a factor of 10 slower th
an the electrons. With moderate cooling to - 20 degrees C and a field
strength around 10 kV/cm, the detectors are homogeneous, i.e. the sign
al response is independent of the interaction location of the gamma-ra
ys. The measured spectra show symmetric full energy peaks with no tail
ing to the low-energy side. Polarization effects at room temperature d
isappear below 0 degrees C. The detectors are well suited for large im
aging arrays in applications such as hard X-ray astrophysics and medic
al imaging. (C) 1998 Elsevier Science B.V. All rights reserved.