REAL-TIME CORRECTION OF CHANCE COINCIDENCE LOSSES IN HIGH-RATE COMPTON SUPPRESSION SPECTROMETRY

Authors
Citation
Gp. Westphal, REAL-TIME CORRECTION OF CHANCE COINCIDENCE LOSSES IN HIGH-RATE COMPTON SUPPRESSION SPECTROMETRY, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 416(2-3), 1998, pp. 536-538
Citations number
8
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
416
Issue
2-3
Year of publication
1998
Pages
536 - 538
Database
ISI
SICI code
0168-9002(1998)416:2-3<536:RCOCCL>2.0.ZU;2-4
Abstract
Chance coincidence losses in a Compton suppression spectrometer are in strumentally corrected by comparing the chance coincidence rate to the counting rate of the Germanium detector. The normally not observable chance coincidence rate is reconstructed from the rates of Germanium d etector and scintillation detector in any auxiliary coincidence unit, after the destruction of true coincidence by sufficiently delaying one of the coincidence partners. (C) 1998 Elsevier Science B.V. All right s reserved.