B. Schwaab et al., TELEMETRY GUIDED PACEMAKER PROGRAMMING - IMPACT OF OUTPUT AMPLITUDE AND THE USE OF LOW-THRESHOLD LEADS ON PROJECTED PACEMAKER LONGEVITY, PACE, 21(11), 1998, pp. 2055-2063
In a prospective study a low threshold screw-in electrode (Medtronic 5
078, group I, n = 9) was compared to a conventional active fixation le
ad (Biotronik Y60BP, group II, n = 9) to investigate whether lower pac
ing thresholds really translate into longer projected service life of
the pacemaker. The leads were implanted in the atrium and were connect
ed to a dual chamber pacing system which included the same ventricular
lead (Medtronic 5024) and the same pulse generator model (Intermedics
294-03) in both groups. Eighteen months after implantation, atrial an
d ventricular pacing thresholds were measured as the charge delivered
per pulse [mu C] at 0.5, 1.0, 1.5, 2.0, and 3.5 V, respectively. For c
hronic output programming in both channels, patients capturing at 0.5
V were set to 1.0 V, those capturing at 1.5 V were permanently program
med to 2.0 V with the double of the charge threshold as the safety mar
gin for pacing (''safety charge''). A combination of atrial and ventri
cular output settings rr as optimal, if it resulted in minimum battery
current drain [mu A] as measured by pacemaker telemetry. In both grou
ps, current consumption [mu A] decreased significantly as output ampli
tude was decreased, exhibiting its lowest value at 1.0 V in either cha
nnel. All ventricular leads could be programmed to the optimum output
amplitude of 1.0 V in groups 1 and 2. As the 2:1 ''safety charge'' val
ues ere almost identical, the ventricular channel essentially contribu
tes the same amount to the battery drain of the pacing system in both
groups. In the atrium, all patients of group 1 could be programmed to
the optimum output amplitude of 1.0 V with an average pulse duration o
f 0.42 +/- 0.15 ms. In group 2, however, ever, all patients had to be
programmed to 2.0 V with a mean pulse width of 0.52 +/- 0.15 ms. With
the atrial and ventricular output being optimized, the average battery
drain of the whole pacing system was 12.19 +/- 0.63 mu A in group 1 v
ersus 14.42 +/- 0.32 mu A in group 2 (P < 0.001). As patients were chr
onically programmed to these output settings, this difference translat
es into a clinically relevant gain in projected pacemaker longevity of
17 months or 18.3 % (121 +/- 4 vs. 104 +/- 2 months; P < 0.001). Thus
, programming a 2:1 safety margin in terms of charge and optimizing th
e output parameters by real-time telemetry of the battery current is a
useful approach to reduce battery current drain. Making the most of m
odern lead technology with a different performance in only one channel
of an otherwise identical DDD pacing system translates into a signifi
cant prolongation of projected pacemaker service life which is of grea
t importance with the increasing awareness of health care expenditures
. The gain in projected longevity is mainly due to the option of reduc
ing the output amplitude which is still significantly beneficial well
below the nominal voltage of the power source.