DEVELOPMENT OF A FAST PAD READOUT SYSTEM FOR THE HADES SHOWER DETECTOR

Citation
A. Balanda et al., DEVELOPMENT OF A FAST PAD READOUT SYSTEM FOR THE HADES SHOWER DETECTOR, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 417(2-3), 1998, pp. 360-370
Citations number
6
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
417
Issue
2-3
Year of publication
1998
Pages
360 - 370
Database
ISI
SICI code
0168-9002(1998)417:2-3<360:DOAFPR>2.0.ZU;2-S
Abstract
The paper presents the readout system for the 18432 channels of the HA DES shower detector. The readout system consists of an analog front-en d ASIC for the readout of the pad multiwire chamber, a resdout control ler board, the so-called readout board for collection and digitization of the analog information from the ASIC and an optical link system fo r the data transfer. The readout system is designed to cope with event rate of 10(5) events/s and a data transfer rate of 1.9 Gbyte/s. Utili zing a multistage fast multiplexing system, the cost per channel is ke pt on the level of $5. Recent results from a test experiment confirm t he intended performance of the readout system. (C) 1998 Elsevier Scien ce B.V. All rights reserved.