We simultaneously employed voltammetry and surface plasmon resonance (
SPR) measurements and monitored electrochemical processes on gold elec
trode. SPR is sensitive not only to the adsorbed layer on the gold ele
ctrode, but also to the dielectric properties of the solution phase. T
herefore, an electrochemically modulated diffusion layer can be detect
ed by SPR. We tested the relationship between the time differential SP
R data and the electrode current for two electrochemical processes and
confirmed the possibility of using electrochemical SPR measurement. (
C) 1998 Elsevier Science S.A. All rights reserved.