NANOLITER TITRATION BASED ON PIEZOELECTRIC DROP-ON-DEMAND TECHNOLOGY AND LASER-INDUCED FLUORESCENCE DETECTION

Citation
E. Litborn et al., NANOLITER TITRATION BASED ON PIEZOELECTRIC DROP-ON-DEMAND TECHNOLOGY AND LASER-INDUCED FLUORESCENCE DETECTION, Analytical chemistry (Washington), 70(22), 1998, pp. 4847-4852
Citations number
12
Categorie Soggetti
Chemistry Analytical
ISSN journal
00032700
Volume
70
Issue
22
Year of publication
1998
Pages
4847 - 4852
Database
ISI
SICI code
0003-2700(1998)70:22<4847:NTBOPD>2.0.ZU;2-P
Abstract
An automated technique for fast acid-base titration of 9-nL sample vol umes has been developed, based on the use of piezoelectrically driven pipets and laser-induced waveguide fluorescence detection. As an examp le, a standard titration of a H2SO4 solution with NaOH was examined by utilizing fluorescein as an: indicator. The technique had reproducibi lity in quantitative determination on the order of 1%. The technique a llows hundreds of individually executed unattended titrations to be pe rformed in a few hours on an area less than 3 cm(2). The lowest amount of acid titrated was similar to 670 fmol.