PYROLYTIC-GRAPHITE X-RAY ECHELON-MONOCHROMATOR

Citation
Ag. Turyanskii et Iv. Pirshin, PYROLYTIC-GRAPHITE X-RAY ECHELON-MONOCHROMATOR, Instruments and experimental techniques (New York), 41(5), 1998, pp. 701-705
Citations number
5
Categorie Soggetti
Instument & Instrumentation",Engineering
ISSN journal
00204412
Volume
41
Issue
5
Year of publication
1998
Pages
701 - 705
Database
ISI
SICI code
0020-4412(1998)41:5<701:PXE>2.0.ZU;2-B
Abstract
A design of an X-ray echelon-monochromator based on semitransparent pl ates of pyrolytic graphite is described. The results of comparative te sts of 0.5- to 1-mm-thick standard pyrographite monochromators, 46- to 180-mu m-thick semitransparent plates, and an echelon-monochromator c omprising three semitransparent plates are presented. An increase in t he peak and integral reflectivity by 55 and 13%, respectively, is achi eved in comparison to the best standard monochromator. The possibility of controlling the spectrum band is shown.