TISSUE EQUIVALENCE CORRECTION FOR SILICON MICRODOSIMETRY DETECTORS INBORON NEUTRON-CAPTURE THERAPY

Citation
Pd. Bradley et Ab. Rosenfeld, TISSUE EQUIVALENCE CORRECTION FOR SILICON MICRODOSIMETRY DETECTORS INBORON NEUTRON-CAPTURE THERAPY, Medical physics, 25(11), 1998, pp. 2220-2225
Citations number
17
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
00942405
Volume
25
Issue
11
Year of publication
1998
Pages
2220 - 2225
Database
ISI
SICI code
0094-2405(1998)25:11<2220:TECFSM>2.0.ZU;2-B
Abstract
Reverse-biased silicon p-n junction arrays have been proposed as micro dosimetry detectors. The tissue equivalence of such detectors in boron neutron capture therapy (BNCT) is discussed. A comparison of the rang e-energy relationships of H, He, C, and Li ions in tissue (ICRU-muscle ) and silicon is given. A simple geometrical scaling (similar to 0.63) of linear dimensions is required to convert microdosimetric energy de position measurements performed in silicon to equivalent deposition in tissue. The Monte Carlo technique is used to examine energy depositio n for two simple geometrical cases applicable to BNCT. (C) 1998 Americ an Association of Physicists in Medicine. [S0094-2405(98)00710-X].