LOW-TEMPERATURE THERMOMETRY IN HIGH MAGNETIC-FIELDS

Citation
Gg. Ihas et al., LOW-TEMPERATURE THERMOMETRY IN HIGH MAGNETIC-FIELDS, Journal of low temperature physics, 113(5-6), 1998, pp. 963-968
Citations number
5
Categorie Soggetti
Physics, Applied
ISSN journal
00222291
Volume
113
Issue
5-6
Year of publication
1998
Pages
963 - 968
Database
ISI
SICI code
0022-2291(1998)113:5-6<963:LTIHM>2.0.ZU;2-8
Abstract
The calibration and reproducibility of va various commercial and homem ade thermometers have been investigated over a temperature range from 20 mK to 4.2 K and magnetic field strength from 0 to 16 Tesla. The per formance of resistance thermometers made of different materials varies widely as a function of magnetic field. One type (Scientific Instrume nts RO600) is found to have a relatively small magneto-resistance, whi ch a's predictable to better than 1% down to 100 mK, making calibratio n in a magnetic field unnecessary for many applications. This was not true of homemade ruthenium oxide and Cernox (Lakeshore Cryotronics). A new type of sensor, thin film doped germanium on a GaAs substrate, wa s also investigated and found to be useful at higher fields and low te mperature.