A new method for the fast screening of cocaine and 6-monoacetylmorphin
e (6-MAM) in a single hair, using gas chromatography/mass spectrometry
(GC/MS), is described. The analyses are conducted in less than 10 min
with minimal sample preparation The novel method combines the Chromat
oProbe direct sample introduction device for intrainjector thermal ext
raction, fast GC separation, a supersonic molecular beam GC/MS interfa
ce and hyper-thermal surface ionization (HSI). The technique has been
successfully employed for the detection of cocaine in as little as a l
-mm section of hair using selected ion monitoring (SIM). Unambiguous f
ull scan mass spectra of cocaine and 6-MAM were obtained on a single h
air for cocaine and heroin users, respectively. HSI was found to be al
most 3 orders of magnitude more selective than electron impact ionizat
ion for cocaine compared with the major hair constituents, with a mini
mum detected concentration of approximately 10 ppb in the SIM mode. Re
sults obtained for 12 drug users showed full qualitative agreement wit
h similar results using rigorous solvent extraction followed by electr
ospray-liquid chromatography/mass spectrometry analysis. However, quan
titative studies showed only partial agreement. No false positives wer
e observed for 10 drug free subjects. This method enables fast drug mo
nitoring along the hair length which permits time correlation studies.
(C) 1998 American Society for Mass Spectrometry.