Kh. Becks et al., COMMISSIONING OF THE DELPHI PIXEL DETECTOR, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 418(1), 1998, pp. 15-21
The DELPHI Vertex Detector has been upgraded for LEP200 with 2 layers
of pixel detectors and 2 layers of ministrips as endcaps, thus coverin
g the polar angular range between 10 degrees and 21 degrees and allowi
ng for standalone pattern recognition. During the 1995/96 shutdown the
first 95 modules were installed and the installation was completed in
May 97 with all 152 modules. Production yields and failures during th
e assembly will be presented. The detector operated at a threshold of
around 9000 electrons resulting in a number of noisy pixels at the 10(
-3) level. After masking these hot pixels in the readout, the remainin
g number of random hits per event is at the 10(-6) level. The observed
resolution of about 100 mu m is close to the expectation for binary r
eadout. The track reconstruction efficiency increased by more than 100
% in the central part of the forward tracker. (C) 1998 Elsevier Scienc
e B.V. All rights reserved.