The use of TL detectors in dosimetry of systemic radiation therapy

Citation
Ac. Aschan et al., The use of TL detectors in dosimetry of systemic radiation therapy, ACTA ONCOL, 38(2), 1999, pp. 189-196
Citations number
32
Categorie Soggetti
Onconogenesis & Cancer Research
Journal title
ACTA ONCOLOGICA
ISSN journal
0284186X → ACNP
Volume
38
Issue
2
Year of publication
1999
Pages
189 - 196
Database
ISI
SICI code
0284-186X(1999)38:2<189:TUOTDI>2.0.ZU;2-7
Abstract
A method for determining absorbed doses to organs in systemic radiation the rapy (SRT) is evaluated. The method, based on thermoluminescent (TL) dosime ters placed on the patient's skin, was Validated and justified through a ph antom study showing that the difference between measured (TL dosimeters in the phantom) and derived (TL method) Values is within 10%. Six radioimmunot herapy (RIT) patients with widespread intraperitoneal pseudomyxoma were als o studied. In dose evaluations, special emphasis was on kidneys. In additio n to the TL method, the absorbed doses to kidneys were calculated using MIR D formalism and a point dose kernel technique. We conclude that in SRT the described TL method can be used to estimate the absorbed doses to those cri tical organs near the body surface within 50% (1 SD).