A model is developed to explore operating conditions and performance tradeo
ffs for high-speed GC using atmospheric-pressure air as carrier gas and a v
acuum pump to draw the carrier gas and injected samples through the system.
The model is based on the rate theory for open. tubular columns and conven
tional equations for gas flow in capillary tubes. The model predicts the ef
fects of column outlet pressure, column length, column diameter, and detect
or dead time on the number of theoretical plates generated in a 30-s analys
is spanning a retention factor range from 0 to 5. The outlet (detector) pre
ssure range considered is 1-100 kPa (0.01-1,0 atm), A 0.1-mm-i.d, column is
found to generate more plates than either larger or smaller diameter colum
ns because of the constraint of using atmospheric pressure at the column in
let. About 25 000 plates are generated with a 2.5-m-long column for outlet
pressures less than similar to 20 kPa, The model is validated with a high-s
peed GC instrument using a cryofocusing inlet system and a photoionization
detector. The number of theoretical plates measured for o-xylene agrees ver
y well with the model predictions for the lowest pressure case, System perf
ormance degrades at higher outlet pressures and with smaller diameter colum
ns because of increased dead time of the detector. Results are considered i
n the context of designing portable GC instruments for ambient VOC analysis
.