Aa. Karyakin et al., The improved potentiometric pH response of electrodes modified with processible polyaniline. Application to glucose biosensor, ANAL COMMUN, 36(4), 1999, pp. 153-156
Processible polyaniline (PCPAn) modified electrodes are characterized by an
advanced potentiometric pH response in comparison to those based on regula
r polyaniline. Glassy carbon electrodes modified with PCPAn by dip-coating
exhibited a fully reversible potentiometric response of approximately 90 mV
pH(-1) over the range pH 3-9. Such significantly higher potentiometric res
ponses of PCPAn modified electrodes compared to existing devices is explain
ed on the basis of the thermodynamics of polyaniline redox reactions. The p
otentiometric biosensor for glucose based on processible polyaniline has be
en developed using a non-aqueous enzymology approach for enzyme immobilizat
ion. In the model solution, which mimics blood serum, the biosensor was use
ful for glucose detection over the concentration range 0.1-30 mM and the ma
ximum response value reached was approximate to 80 mV. The advanced potenti
ometric response of PCPAn modified electrodes provides their application fo
r sensor and biosensor development.