Assessment of serum thyroxine binding capacity-dependent biases in free thyroxine assays

Citation
Nd. Christofides et al., Assessment of serum thyroxine binding capacity-dependent biases in free thyroxine assays, CLIN CHEM, 45(4), 1999, pp. 520-525
Citations number
33
Categorie Soggetti
Medical Research Diagnosis & Treatment
Journal title
CLINICAL CHEMISTRY
ISSN journal
00099147 → ACNP
Volume
45
Issue
4
Year of publication
1999
Pages
520 - 525
Database
ISI
SICI code
0009-9147(199904)45:4<520:AOSTBC>2.0.ZU;2-9
Abstract
Background: Free thyroxine (FT4) assays may exhibit biases that are related to serum T-4 binding capacity (sBC. We describe two tests that can be used to assess the presence and magnitude of sBC-dependent biases in FT4 assays . Methods: We used a direct-equilibrium dialysis FT4 assay as the reference m ethod and compared the results obtained with those of the FT4 assays under investigation, in patient sera having a wide range of sBC, We then compared the expected and observed FT4 results for sera diluted with an inert buffe r. Because serum dilution causes a predictable decrease in sBC, an increasi ngly negative,bias on progressive dilution is indicative of a sBC-dependent bias. Results: The automated FT4 assay investigated. (Vitros FT4) showed no demon strable sBC-dependent bias by either test. Conclusion: These two tests can be used to screen for sBC-dependent biases in FT4 assays. (C) 1999 American Association for Clinical Chemistry.