Objective: This study aimed to investigate the compatibility of the Clarion
1.2 magnet-containing cochlear implant with a 1.5-tesla (T) and 0.3-T magn
etic resonance imager.
Background: Cochlear implants restore functional hearing to patients with s
ensorineural deafness. With the rapidly increasing number of patients with
cochlear implants, there is a need to investigate the implant's magnetic re
sonance imaging (MRI) compatibility.
Methods: The authors tested the potential torque and force on the metallic
components of the implant, heating of the implant and surrounding tissue, u
nintentional output, implant damage, and image distortion. Tests were perfo
rmed in both a 1.5-T and 0.3-T MRI.
Results: The torque experienced by the implant in the 1.5-T MRI (0.19 nm) w
as large enough that it could potentially cause implant movement in some pa
tients. An acceptable amount of torque (0.04 nm) was found in the 0.3-T MRL
. Image distortion occurred in the area directly around the implant with a
radius of up to 60 mm in the 1.5-T MRI and 100 mm in the 0.3-T MRI. In both
MRI units, there was no detectable temperature increase or unintentional o
utput. There was no implant damage except that with worst-case conditions,
the internal magnet was demagnetized by 78.5% with the 1.5-T unit and 3.36%
with the 0.3-T unit.
Conclusions: The authors recommend patients with cochlear implants avoid im
aging in a 1.5-T MRI. The results suggest that the 0.3-T MRI poses little o
r no risks to patients with cochlear implants.