Evaluation of ready-to-use liquid reagents for clinical chemistry in laboratory animals

Citation
H. Tabata et al., Evaluation of ready-to-use liquid reagents for clinical chemistry in laboratory animals, EXP ANIM, 48(2), 1999, pp. 133-136
Citations number
5
Categorie Soggetti
Animal Sciences
Journal title
EXPERIMENTAL ANIMALS
ISSN journal
13411357 → ACNP
Volume
48
Issue
2
Year of publication
1999
Pages
133 - 136
Database
ISI
SICI code
1341-1357(199904)48:2<133:EORLRF>2.0.ZU;2-V
Abstract
We evaluated the ready-to-use liquid reagents for clinical chemistry (6 tes ts), to assess their suitability for use in the toxicology laboratory setti ng. Hitachi 736 automated analyzer was used for the analyses. The evaluatio n included the following studies: Precision, Linearity, Effects of interfer ence substances such as hemolytic hemoglobin, bilirubin, turbidity to the a nalytical values and correlation to the solid reagents, which are prepared each time they are needed. The precision and linearity data were within the reagents' specifications. Results of comparison of the liquid reagents and the solid reagents in analyzing plasma samples of rats, dogs and monkeys w ere generally good except for a bias in results for GOT and GPT, regardless of the animal species tested It is concluded that these types of liquid re agents can be used in clinical pathology examinations in animal studies.