W. Devos et al., Determination of impurities in antique silver objects for authentication by laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS), J ANAL ATOM, 14(4), 1999, pp. 621-626
In addition to visual characteristics, a less manipulable criterion for aut
henticity verification of silver antiques is given by trace and minor eleme
nt patterns in the silver alloy. The analytical method used to analyse prec
ious silver antiques should not visibly damage the object and should enable
the determination of impurities in the ppm-0.5% range. Using laser ablatio
n inductively coupled plasma mass spectrometry (LA-ICP-MS), visible damage
can be restricted to an acceptable minimum. Because most antique silver obj
ects are too large to fit into a normal laser ablation cell, an alternative
cell design was used that allows a direct, virtually non-destructive analy
sis of entire antique silver objects. This cell is placed upon the object t
o be analysed. A micro-amount of the object is then ablated through an aper
ture in the bottom of the cell. The 100 mu m wide craters are almost invisi
ble on an antique silver object. The analytes Zn, Cd, Sn, Sb, Au, Pb and Bi
were measured. Signals were normalized to the Ag signal and silver standar
d materials were used for external calibration. The crater-to-crater repeat
ability of the normalized signals in a homogeneous silver sample was below
10% RSD (Iz = 3) for most elements. Detection limits lie within the sub-ppm
to 2 ppm range. The accuracy was validated with comparative ICP-MS measure
ments after digestion and with X-ray fluorescence (XRF) measurements. The a
nalysis of eight antique silver objects, including one forgery, illustrates
the application of the method.