Mj. Yang et al., Determination of temperature dependence of GaSb absorption edge and its application for transmission thermometry, J APPL PHYS, 85(9), 1999, pp. 6632-6635
Transmission spectra of GaSb have been obtained over a temperature range fr
om 10 to 470 degrees C. Using this information, transmission thermometry is
applied to obtain accurate measurements of sample temperature during molec
ular beam epitaxy growth on GaSb substrates. A GaSb surface reconstruction
transition is determined as a function of Sb flux and substrate temperature
, establishing a laboratory-independent temperature standard. (C) 1999 Amer
ican Institute of Physics. [S0021-8979(99)06707-9].