The Imager for Mars Pathfinder returned over 16,000 high-quality images fro
m the surface of Mars. The camera was well-calibrated in the laboratory, wi
th <5% radiometric uncertainty. The photometric properties of two radiometr
ic targets were also measured with 3% uncertainty. Several data sets acquir
ed during the cruise and on Mars confirm that the system operated nominally
throughout the course of the mission. Image calibration algorithms were de
veloped for landed operations to correct instrumental sources of noise and
to calibrate images relative to observations of the radiometric targets. Th
e uncertainties associated with these algorithms as well as current improve
ments to image calibration are discussed.